• 检验技术 •    

Sysmex XT-2000i血细胞分析仪测定成人网织红细胞相关指标正常参考值

杨可1,孟宪华2,侯静3   

  1. 1. 天津市南开医院
    2. 南开医院检验科
    3. 天津市南开医院检验科
  • 收稿日期:2009-10-14 修回日期:2010-01-14 出版日期:2010-08-15 发布日期:2010-08-15
  • 通讯作者: 杨可

DETECTION ADULT'S REFERENCE RANGE OF RETICULOCYTE PARAMETERS BY SYSMEX XT-2000i HEMATOLOGY ANALYZER

  • Received:2009-10-14 Revised:2010-01-14 Published:2010-08-15 Online:2010-08-15
  • Contact: yang ke

摘要: [摘要] 目的 确定Sysmex XT-2000i血细胞分析仪测定天津市正常成人网织红细胞的参考范围。方法 用Sysmex XT-2000i血细胞分析仪测定102例(男性52例,女性50例)正常成人的网织红细胞六项相关指标,通过统计学方法分析后,确定其正常参考范围。结果 RET%:男性为0.888±0.247;女性为0.850±0.277;男女两组比较无显著性差异(P>0.05);合并后正常值参考范围是0.34-1.40。RET#(×1012/L):男性为 0.04396±0.01243;女性为0.03836±0.01295;两者比较有显著性差异(P<0.05);正常参考范围男性是0.0161-0.0658,女性是0.0125-0.0643。LFR%:男性为98.86±0.85;女性为96.16±0.87;两者比较无显著性差异(P>0.05);合并后正常参考范围是97.3-100。MFR%:男性为0.66±0.64;女性为0.46±0.62;两者比较无显著性差异(P>0.05);合并后正常参考范围是0-1.82。HFR%:男性为0.48±0.59;女性为0.38±0.46;两者比较无显著性差异(P>0.05);合并后正常参考范围是0-1.5。IRF:男性为1.14±0.85;女性为0.84±0.87;两者比较无显著性差异(P>0.05);合并后正常参考范围是0-2.7。结论 网织红细胞相关参数的成人正常参考值随性别﹑仪器﹑地区等因素有所差异,各实验室应根据自己的仪器来确定本室的正常参考范围。

关键词: 网织红细胞, 正常参考值, 血细胞分析仪

Abstract: [Abstract] Objective To establish reference range of reticulocyte counts in healthy adults of Tianjin by Sysmex XT-2000i hematology analyzer. Methods 6 parameters of reticulocyte were evaluated on 102 healthy adults(52men&50women) by Sysmex XT-2000i hematology analyzer and the normal reference range were determined by statistic analysis. Results RET%: male is 0.888±0.247,famale is 0.850±0.277,there is no significant difference between each group(P>0.05),unite reference range of RET% is 0.34-1.40.RET#(×1012/L):male is 0.04396±0.01243,famale is 0.03836±0.01295,there is significant difference between each group(P<0.05),reference range of male is 0.0161-0.0658,famale is 0.0125-0.0643.LFR%:male is 98.86±0.85,famale is 96.16±0.87,there is no significant difference between each group(P>0.05),unite reference range of LFR% is 97.3-100.MFR%:male is 0.66±0.64,famale is 0.46±0.62,there is no significant difference between each group(P>0.05),unite reference range of MFR% is 0-1.82.HFR%:male is 0.48±0.59,famale is 0.48±0.59,there is no significant difference between each group(P>0.05),unite reference range of HFR% is 0-1.5.IRF%:male is 1.14±0.85,famale is 0.84±0.87,there is no significant difference between each group(P>0.05),unite reference range of IRF% is 0-2.7. Conclusion There are differences between reference range of reticulocyte parameters with sex,analyzer,area and so on. Reference range of reticulocyte parameters should be established according to laboratory itself analyzer.

Key words: reticulocyte, reference range, hematology analyzer