天津医药 ›› 2015, Vol. 43 ›› Issue (8): 936-938.doi: 10.11958/j.issn.0253-9896.2015.08.029

• 诊断技术 • 上一篇    下一篇

探讨实时动态血糖监测在2 型糖尿病患者中的准确性

张春风1,葛焕琦1△,谢云2,郑辉1   

  1. 1泰达国际心血管病医院内分泌科(邮编300457);2天津医科大学代谢病医院神经内科
  • 收稿日期:2014-12-17 修回日期:2015-03-26 出版日期:2015-08-15 发布日期:2015-08-15
  • 通讯作者: 葛焕琦 E-mail:gehuanqi@163.com
  • 作者简介:张春风(1989),女,主治医师,硕士,主要从事糖尿病及相关并发症发病机制研究

Investigate the accuracy of real-time continuous glucose monitoring system in the type 2 diabetes subjects

ZHANG Chunfeng1, GE Huanqi1△, XIE Yun2, ZHENG Hui1   

  1. 1 Department of Endocrinology, TEDA International Cardiovascular Disease Hospital, Tianjin 300457, China 2 Department of Neurology, Metabolic Disease Hospital of Tianjin Medical University
  • Received:2014-12-17 Revised:2015-03-26 Published:2015-08-15 Online:2015-08-15

摘要: 目的探讨实时动态血糖监测(RT-CGM)在2 型糖尿病(T2DM)患者中的准确性。方法选取2013 年5 月—2014 年8 月于我院住院治疗的T2DM 患者318 例作为研究对象。所有受试者佩戴RT-CGM 检测仪,每日输入至少4 个指端血糖进行校正。依指端血糖水平将受试者分为3 组:≤6.99 mmol/L 组、7.0~10.0 mmol/L 组和≥10.01 mmol/L 组,分析指端血糖与探头血糖的相关性,并应用Clarke 错误栅格分析评价RT-CGM 的准确性。结果(1)本研究共得到2 815 对指端血糖-探头血糖数据,探头血糖与指端血糖呈正相关(r=0.847),对指端血糖水平分组后分析显示:两者的相关性随血糖值的升高而升高(r 分别为0.457、0.648 和0.852,P<0.001)。(2)Clarke 错误栅格分析显示:RT-CGM 值的临床准确率为76.69%(A 区),可接受率为20.28%(B 区),临床不可接受血糖比例为3.03%(C~E 区)。指端血糖≤6.99 mmol/L 时的临床准确率低于7.0~10.0 和≥10.01 mmol/L 时(73.12% vs 78.63% & 79.28%,P< 0.05),而后2 组水平的临床准确率差异无统计学意义。结论RT-CGM 具有良好的准确性,尤其是在高血糖时其准确性更稳定。

关键词: 糖尿病, 2 型, 实时动态血糖监测, 准确性, 错误栅格分析

Abstract: Objective To investigate the accuracy of real-time continuous glucose monitoring(RT-CGM)system in the type 2 diabetes mellitus(T2DM). Methods A total of 318 subjects hospitalized between 2013 May to 2014 August were recruited. Each subject received a RT-CGM system. In order to calibrate the interstitial glucose level, finger-stick blood glucose were measured at least four times every day. Patients were divided into three groups according to their fingerstick blood glucose level: ≤6.99 mmol/L, 7.0-10.0 mmol/L and ≥10.01 mmol/L. Pearson correlation was used to analyze the relationship between the finger-stick glucose and interstitial glucose gained from RT-CGM. Clarke error grid analysis was used to analyze the accuracy of RT-CGM. Results (1)A total of 2, 815 glucose meter values from finger-stick were com⁃ paired with glucose from RT-CGM. A positive relationship(r=0.847)were found between CGM and finger-stick blood glu⁃ cose. The correlation was closer with higher finger-stick blood glucose levels(r=0.457, 0.648 and 0.852 respectively in ≤ 6.99 mmol/L group, 7.0-10.0 mmol/L group and ≥10.01 mmol/L group. P<0.001).(2)The Clarke error grid analysis re⁃ vealed that 76.69% of the readings from RT-CGM were clinical perfect (zone A), 20.28% were acceptable (zone B), 3.03% were unacceptable (zone C-E). Clinical perfect rate was lower when the finger-stick glucose ≤6.99 mmol/L , compared with that when it is between 7.0-10.0 or ≥10.01 mmol/L group(73.12% vs 78.63% & 79.28%,P<0.05), there was no significant difference between group of 7.0-10.0 mmol/L with group ≥10.01 mmol/L (P>0.05). Conclusion RT-CGM provides accu⁃ rate blood glucose values to guide the treatment of diabetes, and the accuracy is more reliable in high glucose environment.

Key words: diabetes mellitus, type 2, real-time continuous glucose monitoring, accuracy, error grid analysis